CALYS 75,Calibrator
A complete tool able to measure and generate all manufacturing process signals: voltage, current, resistance, temperature by resistive probes and thermocouples, frequency, counting and measurement and simulation on dry contacts.
Dual display continuously shows the measurement value, the emitted value, and indicates the date, time and external temperature.
Drop-down menus and online help available.
Features the ultra simple "Easyconnect" system, which allows you to connect sockets, wires, banana plugs and security connectors, as well as flat thermocouple connectors.
Emission of ramps and steps to test linearity and transmitters with a delay function.
Graphic display with date and time, backlit display with contrast settings Display resolution may be selected with 1, 2 or 3 digits after the point.
Measurement statistics displayed: Minimum, Maximum and Average.
10,000 values may be stored.
Shockproof sheath.
Integrated valve test.
Scales quadratic signals.
Compatible with HART transmitters.
USB connection for downloading.
Supplied with measuring cables, strap, stand for desktop use and fitted with European plug.
VOLTAGE
Measurement 100 mV 0.001 mV 0.013 % + 2 μV
1 V 0.01 μV 0.013 % + 20 μV
10 V 0.1 μV 0.013 % + 200 μV
50 V 01 mV 0.013 % + 2 mV
Emission 100 mV 0.001 mV 0.013 % + 2 μV
2 V 0.01 μV 0.013 % + 20 μV
50 V 0.1 μV 0.013 % + 2 mV
CURRENT
Measurement 50 mA 0.001 mA 0.018 % + 2 μA
Emission 24 mA 0.001 mA 0.018 % + 2 μA
RESISTANCE
Measurement 400 Ω 0.001 Ω 0.012 % + 10 mΩ
4000 Ω 0.01 Ω 0.012 % + 100 mΩ
Emission 400 Ω 0.001 Ω 0.017 % + 20 mΩ
4000 Ω 0.01 Ω 0.017 % + 200 mΩ
FREQUENCY
Measurement 20 KHz 0.01 Hz 0.005 %
Emission 10 Hz 0.01 Hz 0.005 %
10 Hz 10 Hz 0.005 %
PT100
Measurement -210 → +850 °C 0.01 °C 0.012 % + 0.05 °C
Emission -210 → +850°C 0.02 °C 0.017 % + 0.08 °C
Type K Thermocouple
Measurement -250 → +1372 °C 0.05 °C 0...